Integra LifeSciences Holdings Corporation · NASDAQ:IART
Integra LifeSciences — Investor Presentation (2015)
35 slides from Integra LifeSciences's investor presentation, 2015
Slide 1: Codman Hakim Programmable Valve (CHPV)
Codman Hakim Programmable Valve (CHPV) ® Company Confidential – Not for Distribution 1
Slide 2: CODMAN Neuro Programmable Valve History
CODMAN Neuro Programmable Valve History 2013 SIPHONGUARD CODMAN ® Device CERTAS PLUS™ 2002 2007 Programmable Valve CODMAN HAKIM 2006 Silicone Programmable Valve BACTISEAL Stator fit housing 1992 1998 Improvement redesign 1999 2003 2007 Second Cam redesign VPV Generation Programmer Company Confidenti
Slide 3: Adjustable Valves
Adjustable Valves • Valve Mechanism allows the change operating pressure noninvasively • Codman Valve – Codman Programmable Valve – 18 pressure settings Company Confidential – Not for Distribution 3
Slide 4: How does CHPV work?
4 How does CHPV work ? Spring Spiral Cam Ruby Ball & Seat Base Plate CODMAN HAKIM Programmable Valve (CHPV) Company Confidential – Not for Distribution 4
Slide 5: The CHPV technology
5 The CHPV technology Codman uses a cantilever spring design in concert with a perfectly matched ruby ball and valve seat. • Maintains a precise constant opening pressure over years of service. • Silicone valve performance degrades over time. • Tight operating pressure tolerance. Company Confidentia
Slide 6: The key features of CHPV
The key features of CHPV • Eighteen (18) Distinct Pressure Settings • 30 to 200 mm H20 • Precise Pressure Bandwidths • 8 configurations • Available unitized with anti siphon device: SIPHONGUARD Device Company Confidential – Not for Distribution 6 6
Slide 7: The key features of CHPV
The key features of CHPV • MR conditional up to 3 Tesla • Smallest MRI artifact in the market. Max signal void area = 2439mm² • Minimal magnetic influence (min 80 Gauss): electronic devices such as Iphone and Ipad can be safely used by patients when used as intended • Shunt placement site does not a
Slide 8: The CHPV: The most precise valve in the market
8 The CHPV: The most precise valve in the market 225 200 • No siphoning effect 175 145 150 • 18 settings that range from 30 -200mmH2O. 115 125 • Enables the physicians to make small and 100 accurate adjustments 80 of 10mmH2O 75 Increments throughout the treatment course 40 50 of their patients. 20 2
Slide 9: SiphonGuard Siphon / Flow Control
SiphonGuard® Siphon / Flow Control A unique device designed to reduce the risk of CSF overdrainage complications. • Dual-pathway system – Prevents excessive flow while allowing constant drainage • Stable mechanical system – Remains reliably sensitive to differences between normal and excessive flow
Slide 10: Different valve configurations
10 Different valve configurations CHPV with SG Certas Plus + with SG SG standalone Company Confidential – Not for Distribution 10
Slide 11: SIPHONGUARD: Dual Pathways
SIPHONGUARD: Dual Pathways • The primary pathway remains open at physiological flow rates ≤40 ml/hr (twice the normal average CSF production rate in adults) • The primary pathway closes at a flow rate of about 100 ml/hr and CSF then flows through (and only through) the secondary pathway, at 10% of t
Slide 12: SiphonGuard - Simplified
SiphonGuard - Simplified TM Primary Pathway Secondary Pathway Proximal Company Confidential – Not for Distribution 12
Slide 13: X-Ray Technique and X-Ray Overlay Tool
X-Ray Technique and X-Ray Overlay Tool Valve / Plate Orientation Company Confidential – Not for Distribution 13 13
Slide 14: X-Ray Valve Information
X-Ray Valve Information • The white marker on the valve (1) indicates the right hand side of the valve – The marker will not be seen on the x-ray if it is positioned on a 30° angle or more; the valve cannot be programmed if the angle is 45° or more • The pressure indicator on the white ring (2) deno
Slide 15: Codman VPV
Codman VPV • Overview • Tips for use Company Confidential – Not for Distribution 15
Slide 16: The CHPV programming
16 The CHPV programming • Not painful • Accurate • Adaptability to a variety of clinical situations Company Confidential – Not for Distribution Data Source: Codman Data on File 16
Slide 17: Company Confidential – Not for Distribution
Company Confidential – Not for Distribution 17
Slide 18: VPV
VPV • Sensitive Sonar technology • Listens to the valve • Counts the steps • Confirms the adjustment Company Confidential – Not for Distribution 18
Slide 19: A new programmer
A new programmer Transmitter head A. Start button B. Direction of flow arrow C. Cord D. Central rod (Acoustic device) E. Feet Company Confidential – Not for Distribution 19
Slide 20: Technique tips
Technique tips • Be aware of noise levels • Transmitter head positioning Programming Zone: Equal to existing programmer Listening Zone: Center coupling rod over cam Company Confidential – Not for Distribution 20
Slide 21: Technique tips
Technique tips • A pea-sized amount of gel on the central rod • Might also be applied on the patient’s head over the valve mechanism. • Avoid gel contact with programmers’ probes. Programming Zone: Equal to existing programmer Listening Zone: Center coupling rod over cam Company Confidential – Not f
Slide 22: A new programmer
A new programmer Two modes • Packaged valve • Implanted valve Company Confidential – Not for Distribution 22
Slide 23: A new programmer
A new programmer Packaged valve mode: allows the adjustment of the valve before the implantation, through the sterile blister. When in this modality, the acoustic sensor is de-activated, and the programmer is equal to model 82-3190R. Implanted valve mode: allows the adjustment of the valve noninvasi
Slide 24: VPV
VPV • Confirms adjustment – It needs to count the clicks • Compatible with all CHPV • Avoids “false positives” Company Confidential – Not for Distribution 24
Slide 25: VPV system
VPV system Is the machine in working order? • Previous “Adjustment complete” message, and no system-error messages indicates it is OK • Defect would give a system-error message or would mean you would never get an “adjustment complete” message Company Confidential – Not for Distribution 25
Slide 26: VPV System
VPV System • Quick Check – centre rod on the transmitter head moves in and out with GENTLE pressure. If not, then it may have been damaged recently. Company Confidential – Not for Distribution 26
Slide 27: VPV performance
VPV performance • Circumstances that could affect VPV performance – Position of the valve • Close to ear = difficult to get VPV transmitter parallel to valve base • Solution = attempt to keep parallel orientation even if it means only some of the transmitter feet contact the patient • Transmitter he
Slide 28: VPV performance
VPV performance • Contour of the skull – Esp cranial defects can alter the orientation of the transmitter – Need the transmitter parallel rather than rely 100% on the feet to produce the correct parallel orientation – If the valve is prominent, avoid positioning the head so that the center rod fully
Slide 29: VPV performance
VPV performance • Skin thickness and scarring – In testing skin thickness was not an issue – Scar tissue can be a problem • Abundance of hair – Better to flatten and part the hair – The closer the rod is to contacting the skin the better – If hair causes air gap between center rod and skin, fill the
Slide 30: VPV performance
VPV performance • Noises at time of adjustment – If patient coughs, talks, snores, moves or causes a vibration during the adjustment this will be interpreted as an unexpected event – “repeat adjustment” – If noise occurs, then repeat with elimination of noise Company Confidential – Not for Distribut
Slide 31: Note:
Note: • VPV is more sensitive to transmitter/valve position. Positioning the centre rod directly over the cam mechanism should be the main priority. • This can be facilitated by marking the skin above the cam with skin marker Company Confidential – Not for Distribution 31
Slide 32: Users Tips
Users Tips • Arrow on transmitter head must be in line with axis of flow • Centre rod must be above the CHPV ideally above the moveable cam • Transmitter head should be parallel to the valve- valve not rotated on the axis of flow Company Confidential – Not for Distribution 32
Slide 33: Users Tips
Users Tips • Properly position the center rod on the valve mechanism – Possibly because • Previous success with 82-3190 with familiar method of positioning transmitter head relative to valve ( same experienced users now get “repeat adjustment” message sometimes because too proximal or distal to cam
Slide 34: Manufacturing
Manufacturing Le Locle, Switzerland Clean Room Company Confidential – Not for Distribution 34
Slide 35: Indications and contact information
Indications The Codman® Hakim® Programmable Valves are implantable devices that provide constant intraventricular pressure and drainage of CSF for the management of hydrocephalus. Contraindications The Codman® Hakim® Programmable Unitized Valve Systems are not recommended for atrial placement. Use t