Slide 1: X-am 5100 Monitor Training
© Google, 2007 X-am 5100 MONITOR TRAINING Marketing, December 2012
Slide 2: Disclaimer
DRÄGER X-AM 5100 DISCLAIMER This visual presentation is intended to help the user understand and correctly use the Dräger X-am 5100 instruments. It is not intended to be a substitute for the complete instruction manual supplied with the instrument. IT IS THE USER’S RESPONSIBILITY TO READ AND UNDERST
Slide 3: General Information
Drager General Information Drager. Technology for Life*
Slide 4: X-am 5100 Description
DRÄGER X-AM 5100 X-am 5100 DESCRIPTION The Dräger X-am 5100 Monitor is a single gas measuring and alarm device designed for use in personnel monitoring applications. Use to detect: Hydrogen Chloride Hydrogen Fluoride Hydrazine Hydrogen Peroxide DrägerSensor® patented technology 4 | 48 X-am 5100
Slide 5: Enclosure
DRÄGER X-AM 5100 ENCLOSURE • Size: 1.89 x 5.12 x 1.89 in. (48 x 130 x 61 mm) • Weight: ~7.8 to 8.8oz (220g to 250g) • IP 54 approved - Dust & water resistant • Over-molded rubber boot offers additional protection • RFI Protected: CE mark electromagnetic compatibility (89/336/EEC) • Temp Range - cont
Slide 6: X-am 5100 Description: Datalogging
DRÄGER X-AM 5100 X-am 5100 DESCRIPTION Datalogging is standard in X-am 5100 • >1,000 hours recording at 1 minute intervals • Events are also logged 6 | 48 X-am 5100 | Training| 04.2012
Slide 7: Basics – Components on Front
DRÄGER X-AM 5100 BASICS – COMPONENTS ON FRONT Visual alarm Visual alarm XS sensor Horn > 90 db Dot Matrix display +M – Button OKon – Button Navigation Switch on Acknowledge 7 | 48 X-am 5100 | Training| 04.2012
Slide 8: Instrument Components – Rear
DRÄGER X-AM 5100 INSTRUMENT COMPONENTS - REAR IR Interface Grip Clip Approval Label Battery Pack Battery Capture Screw Battery Charge Contacts - NiMHy 8 | 48 X-am 5100 | Training| 04.2012
Slide 9: X-am 5100 Battery Options
DRÄGER X-AM 5100 X-am 5100 BATTERY OPTIONS • X-am 5100 Units come standard with a NiMH battery pack, p/n 8318704 or ‘AA’ alkaline battery pack using 2 cells • Optional high-capacity NiMH pack is available, p/n 8322244 9 | 48 X-am 5100 | Training| 04.2012
Slide 10: Battery Life
DRÄGER X-AM 5100 BATTERY LIFE • X-am 5100 Standard Operating Times - Diffusion Standard NiMH Battery >120 hrs High-Capacity NiMH Battery >180 hrs “AA” Alkaline Batteries >120 hrs • Operating times are shortened by; Excessive Alarms Colder temperatures (less than 50 ºF / 15 ºC) 10 | 48 X-am 5100
Slide 11: Charging Cradle
DRÄGER X-AM 5100 CHARGING CRADLE • The Charging Cradle provides a full, managed charge to the NiMHy battery pack in < 4 hours from complete discharge • Up to 20 modules can be ganged together for use with one 8315805 Multi- Charger Power supply • Up to 5 modules can be ganged together for use with 8
Slide 12: X-am 5100 Operation
Drager X-am 5100 Operation Drager. Technology for Life*
Slide 13: Button Operation
DRÄGER X-AM 5100 2 BUTTON OPERATION Operation controlled by a 2-button key pad “OK / On”Button Press and hold to turn monitor ON “+M” Button Press and hold “+M” when monitor is off to verify installed sensors and battery charge Press and hold both “OK” and “+M” to turn monitor OFF 13 | 48 X-am
Slide 14: Switching-On
DRÄGER X-AM 5100 SWITCHING-ON DRÄGER X-AM 5100 • Press and hold the “OK”-key for 3 seconds • Display counts down 3-2-1 • LEDs Flash, the audible alarm sounds and the vibrating alarm pulsates • The self-test begins automatically and shows the following: -The installed software version -The alarm leve
Slide 15: Switching Off
DRÄGER X-AM 5100 SWITCHING OFF DRÄGER X-AM 5100 • Press and hold the “+” and “OK" keys at the same time for more than 3 sec. • The LEDs will flash and a beep will sound. • The instrument is switched off 15 | 48 X-am 5100 | Training| 04.2012
Slide 16: Display
Drager Drager. Technology for Life*
Slide 17: Display / Info Mode
DRÄGER X-AM 5100 DISPLAY / INFO MODE Display • Large, easy to read numbers • Backlight turns on when a key is pressed or when in an alarm condition • Shows gas, units, reading, and battery level • Indicates alarm conditions or faults • Info Mode • Press the “+M” key when the monitor is OFF and the i
Slide 18: Display Layout
DRÄGER X-AM 5100 DISPLAY LAYOUT The X-am 5100 display is divided into 3 sections 1. Gas Type 2. Gas Values and Concentration units 3. Special Symbols 1 2 3 18 | 48 X-am 5100 | Training| 04.2012
Slide 19: Display – Special Symbols
DRÄGER X-AM 5100 DISPLAY – SPECIAL SYMBOLS Special Symbols provide a quick message about the instrument condition Error Notice Peak Battery 19 | 48 X-am 5100 | Training| 04.2012
Slide 20: Display Illumination
DRÄGER X-AM 5100 DISPLAY ILLUMINATION • Press any key and the display illumination is switched on for about 10 sec. • In an alarm situation, the display illumination is automatically switched on 20 | 48 X-am 5100 | Training| 04.2012
Slide 21: Display – Special Symbols
DRÄGER X-AM 5100 DISPLAY – SPECIAL SYMBOLS The Notice symbol indicates there is something that the user should be aware of. The instrument is still usable!! Check / Service instrument at the end of the shift 1 2 3 21 | 48 X-am 5100 | Training| 04.2012
Slide 22: Alarms
Drager Drager. Technology for Life*
Slide 23: Alarm Indications
DRÄGER X-AM 5100 ALARM INDICATIONS Four Alarm indicators are standard with every model to ensure the customer is aware of potential hazards: A loud audible horn - >90 dB @ 1ft. (30 cm) Three bright flashing, red LED’s An Alarm level indication on the LCD Display A powerful vibrating alarm Distin
Slide 24: A1 Gas Concentration Alarm
DRÄGER X-AM 5100 A1 GAS CONCENTRATION ALARM • Audible-, visual- and vibrating alarms repeat once per second • The display character “A1” will alternate with the concentration in the display • An A1 alarm, the audible and vibrating alarms can be acknowledged by pushing the A1 “OK ”-key. 24 | 48 X-am
Slide 25: A2 Gas Concentration Alarm
DRÄGER X-AM 5100 A2 GAS CONCENTRATION ALARM • Double audible-, visual- and vibrating alarms repeat twice per second 20 • The display character “A2” will alternate with the concentration in the display • The audible, visual and vibrating alarms can NOT be acknowledged (silenced) in A2 “A2” alarm Foll
Slide 26: A1 Low Battery Alarm
DRÄGER X-AM 5100 A1 LOW BATTERY ALARM • Audible-, visual- and vibrating alarms repeat once per second • Icon flashes • The alarm can be acknowledged with “OK”-key This is activated when the battery has less than 10 minutes of operation – please change the batteries or charge the instrument. 26 | 48
Slide 27: A2 Low Battery Alarm
DRÄGER X-AM 5100 A2 LOW BATTERY ALARM • Audible-, visual- and vibrating alarms repeating two times per second • flashes • The instrument will automatically shut down in about 10 seconds • Instrument switches off • The alarm can NOT be acknowledged 27 | 48 X-am 5100 | Training| 04.2012
Slide 28: Error Alarm
DRÄGER X-AM 5100 ERROR ALARM • Audible-, visual- and vibrating alarms triple repeat each second -- • The icon appears in the display • These alarms may be silenced with the “OK” - key, but the gas display will still indicate a fault An error is an indication that something needs to be looked at in t
Slide 29: Over Range
DRÄGER X-AM 5100 OVER RANGE • Audible-, visual- and vibrating alarms repeat two times a second • and A2 appear alternately • The alarm can NOT be acknowledged • Sensor reading has exceeded the measuring range If the measuring range is exceeded, the following display is shown instead of measured valu
Slide 30: Bump Test
Drager Bump Test Drager. Technology for Life*
Slide 31: Bump Test
DRÄGER X-AM 5100 BUMP TEST • Switch on the instrument. • Prepare a test gas source (e.g., cylinder, permeation oven); the volume flow must be 0.5 L/min and the gas concentration must be higher than the alarm threshold concentration to be tested. • Fit the calibration adapter (68 06 291) to the senso
Slide 32: Charging & Maintenance
Drager Charging & Maintenance Drager. Technology for Life*
Slide 33: Charging the Battery Pack
DRÄGER X-AM 5100 CHARGING THE BATTERY PACK • Slide the instrument into the charging module • Connect the charging module with a single- or multi-charger • The instrument is charging • Red LED blinks once per second -> instrument is charging • Red LED lights stay on -> instrument is fully charged • R
Slide 34: Battery Case
DRÄGER X-AM 5100 BATTERY CASE • Loosen the screw with the help of a 2.0 mm Allen key, included with the monitor • Remove the battery pack by sliding it down out of the housing • Install the replacement battery and tighten the screw Do not charge the battery in explosion hazard areas. Alkaline-Batter
Slide 35: Environmental Conditions
Drager Environmental Conditions Drager. Technology for Life*
Slide 36: Environmental Conditions – Temperature
DRÄGER X-AM 5100 ENVIRONMENTAL CONDITIONS - TEMPERATURE Temperature • The X-am 5100 is designed for continuous operation between -4 to 122 F (-20 to 50 C) with NiMH battery • Use outside of these parameters is possible, consult X-am 5100 and Sensor specifications for details • Rapid or drastic chang
Slide 37: Environmental Conditions – Humidity
DRÄGER X-AM 5100 ENVIRONMENTAL CONDITIONS - HUMIDITY Humidity • The X-am 5100 is designed for continuous operation between 5 to 95 %RH • Use outside of these parameters is possible, consult X-am 5100 and sensor specifications for details • Rapid or drastic changes in humidity may cause slight fluctu
Slide 38: Environmental Conditions – Pressure
DRÄGER X-AM 5100 ENVIRONMENTAL CONDITIONS - PRESSURE Pressure • The X-am 5100 is designed for continuous operation between 20.7 to 38.4 in. Hg (700 to 1300 mbar) • Rapid or drastic changes in pressure (entering pressurized vessels or large changes in elevation) may cause slight fluctuations of the s
Slide 39: Environmental Conditions – Dust & Mist
DRÄGER X-AM 5100 ENVIRONMENTAL CONDITIONS – DUST & MIST Dust & Mist • If dust or mist (including splashing water) is allowed to accumulate on the sensors, it can prevent gas from entering the sensors • Sensor position does not affect the ability to detect gas or sensor specifications • If the X-am 5
Slide 40: Review of Basics
Drager Review of Basics Drager. Technology for Life*
Slide 41: Quick Review
DRÄGER X-AM 5100 QUICK REVIEW • To turn the X-am 5100 “ON” press and hold the “OK” button for a 3-2-1 countdown • If there is an alarm during use, read the display for information icons and gas values • Take the appropriate safety precautions • To turn the X-am 5100 “OFF”, press and hold both the “O
Slide 42: Thank You!!