Slide 1: Ablation Concepts Using the EPT-1000TC Generator
Ablation Concepts Using the EPT-1000TC™ Generator Copyright ©2003 by Boston Scientific or its affiliates. All rights reserved. EPT-10038_09/01
Slide 2: Ablation Concepts: What is Radio Frequency (RF) energy?
Ablation Concepts • What is Radio Frequency (RF) energy? A form of alternating current (AC) Similar to a wall socket but at a much higher frequency (500,000 Hz vs. 60 Hz) Ideal energy form for ablation because it heats tissue but does not cause muscle stimulation Copyright ©2003 by Boston Scie
Slide 3: What is the goal of an RF ablation?
What is the goal of an RF ablation? • To turn electrically active tissue into electrically unexcitable tissue Generate sufficient heat to cause thermal injury When tissue temperature is elevated beyond 45oC cells are vulnerable to irreversible damage Copyright ©2003 by Boston Scientific or its a
Slide 4: Ablation Concepts: Heat is a form of energy
Ablation Concepts • Heat is a form of energy Z = impedance (sometimes referred to as resistance or R) V = voltage (a potential created between opposite charges) I = current (the flow of electrical charge) Energy = (power) (time). For ablation, we want to deliver lots of energy quickly -- but
Slide 5: Ablation Concepts: Applying power through the generator
Ablation Concepts • When we apply power (watts) through the generator: Electrons will flow easily within the catheter and cable to the ground pad (very low impedance along metal structures) Between the catheter tip and the ground pad, charge build up causes an electric potential or voltage Thi
Slide 6: Ablation Concepts: Critical Circuit
Ablation Concepts • Critical Circuit Copyright ©2003 by Boston Scientific or its affiliates. All rights reserved. EPT-10038_09/01
Slide 7: Energy Flow During Ablation
Ablation Concepts Energy Flow During Ablation • Ratio of RF energy Tissue vs. Blood Determined by tip contact No contact with tissue means all RF energy is going to blood = no tissue heating = no lesion • Good contact angle is critical to lesion growth Copyright ©2003 by Boston Scientific or i
Slide 8: Ablation Concepts: Energy flows to tissue and blood
Ablation Concepts Energy flows to both tissue and blood TIP BLOOD BLOOD TISSUE TISSUE Copyright ©2003 by Boston Scientific or its affiliates. All rights reserved. EPT-10038_09/01
Slide 9: Ablation Concepts: Lesion formation
Ablation Concepts Lesion formation • Lesion size is determined by many factors: – Tip electrode length (4mm or >) – Tip electrode diameter (Fr size) – Power (watts) available to ablate Copyright ©2003 by Boston Scientific or its affiliates. All rights reserved. EPT-10038_09/01
Slide 10: Lesion Formation
Lesion Formation Copyright ©2003 by Boston Scientific or its affiliates. All rights reserved. EPT-10038_09/01
Slide 11: Lesion Formation
Lesion Formation • Two major forces at tissue/electrode interface: Conduction = the transfer of heat from shallow tissue to deeper tissue and from tissue to electrode tip Convection = the cooling effects from the local blood pool • Lesion size is determined mostly by how much heat is transferred
Slide 12: Lesion Formation
Lesion Formation BLOOD TIP BLOOD LESION HOTTEST SPOT TISSUE Copyright ©2003 by Boston Scientific or its affiliates. All rights reserved. EPT-10038_09/01
Slide 13: Coagulum Formation
Coagulum Formation • Coagulum is a “film of coagulated blood” covering the electrode tip • This high temperature causes the blood serum proteins to denature & coagulum forms on the tip • A sudden sharp rise in impedance follows excessive coagulum formation • In ANY system, there can be up to a 20 de
Slide 14: Thrombus Formation
Thrombus Formation • Thrombus may be used to describe coagulum but they are 2 different events • Coagulum is generally used to describe a “film” covering the electrode tip • Thrombus is also a clotting of blood but is found along the catheter body and can result from: Using too large of an introdu
Slide 15: Charring and Popping
Charring and Popping • Charring occurs when voltage levels get too high It is not good because char could break off as embolic debris in bloodstream • Popping occurs when tissue temperatures exceed 100 degrees C and steam is produced within tissue Clinical ramifications of emboli, perforations •