Nemaura Medical — Technical Presentation (December 2018)

11 slides from Nemaura Medical's technical presentation, Dec 18, 2018

  1. Slide 1: Better Diagnostics for Life: Clinical Presentation sugarBEAT

    NASDAQ: NMRD Better Diagnostics for Life Clinical Presentation sugarBEAT® 18th December 2018

  2. Slide 2: A Prospective Single Centre Evaluation of the Accuracy and Safety of sugarBEAT

    A Prospective Single Centre Evaluation of the Accuracy and safety of the sugarBEAT® Non-invasive Continuous Glucose Monitor (CGM) System Dr T Rahman, MBBS, PhD Clinical research Manager, Nemaura Medical Inc., 2

  3. Slide 3: Study Objective

    Study Objective A EU based study to determine the safety and accuracy of the sugarBEAT® CGM for FDA De Novo submission 3

  4. Slide 4: Study Design

    Study Design ❖Prospective single arm, single centre study ❖Accuracy and safety assessed over 7 consecutive wear days, consisting of 3 non-consecutive in-clinic visits, and 4 home wear days. ❖Venous blood samples used as reference for in-clinic portion of study using Architect c8000 ❖Abbott Freestyle

  5. Slide 5: Study Methods

    Study Methods ❖>25 subjects enrolled for screening at each of 3 stages ❖25 subjects selected after screening in each stage, with approximately equal split between Type 1 and Type 2 ❖No subjects lost due to drop out ❖All 25 subjects completed 2 days home study and 3 days in- clinic study ❖12 of each

  6. Slide 6: Accuracy vs Glucose Range In-clinic 1-point calibration

    Accuracy vs Glucose Range In-clinic 1-point calibration The proportion of agreement is 76.098% (with 95% confidence interval from 0.75382 to 0.76814). Table 1. shows the proportion of agreement broken down for different glucose and accuracy ranges and MARD / MAD. MARD +/-20% or +/- 20mg/dL (76% of p

  7. Slide 7: Accuracy vs Glucose Range In-clinic 1-point calibration

    Accuracy vs Glucose Range In-clinic 1-point calibration MAD: Mean Absolute Deviation MARD: Mean Absolute Relative Deviation The performance evaluation included the proportion of the CGM system values that are within ±10 to >40% of relative difference of reference value at glucose levels >80 mg/dL an

  8. Slide 8: Accuracy vs Glucose Range In-clinic 2-point Calibration

    Accuracy vs Glucose Range In-clinic 2-point Calibration The proportion of agreement is 78.33% (with 95% confidence interval from 0.77646 to 0.79032). Table 1. shows the proportion of agreement broken down for different glucose and accuracy ranges and MARD / MAD. MARD +/- 20% or +/- 20mg/dL (78.7% of

  9. Slide 9: Accuracy vs Glucose Range In-clinic 2-point Calibration

    Accuracy vs Glucose Range In-clinic 2-point Calibration MAD: Mean Absolute Deviation MARD: Mean Absolute Relative Deviation Note 1: Number of paired points is different/higher compared 1-point calibration in light of better accuracy leading to more data points falling within the 40-400mg/dL range Th

  10. Slide 10: Safety Evaluation

    Safety Evaluation Previous studies: >100 Patient days wear time – No major adverse events or skin irritation reported. >525 Patient days wear time – No major adverse events or skin irritation reported. Current study: >250 Patient days to date with no device related adverse events or skin irritation

  11. Slide 11: Conclusion

    Conclusion The analysis indicates the in-clinic portion of the study shows accuracy levels, as measured using overall nominal MARD, as comparable to previously reported data and within the targets for the intended indications. 11